LACTATE DEHYDROGENASE-LDH
Intended Use:
Test for estimation of LDH activity in serum / plasma using UV-Kinetic method.
LDH Test Procedure
LDH Programming Video
Possible Errors in LDH and Solutions
Features:
- Method Specific To Lactate Dehydrogenase
- Liquid Stable Reagents With Long Shelf Life
- Suitable For Fully Automated Chemistry Analyzer
- Fast Result
- High Linearity
Specification
Methodology | P L UV-Kinetic |
Reaction Type | Kinetic |
Sample Type | Human Serum or Plasma |
Number Of Reagents | Two |
R1:R2 Mixing Ratio | 4:1 |
Primary Wavelength | 340 nm |
Secondary Wavelength | NA |
Flow Cell Temperature | 37 0C |
Blank Settings | Distilled Water |
Blank Absorbance Limit | >1.00 |
Reagent Volume Required Per Test | 1000 µl |
Sample Volume Required Per Test | 20 µL |
Read Time | 120 Sec |
Incubation Temperature | 37 0C |
Standard Available in the Kit | Yes |
Standard Concentration | NA |
Factor | 8109 |
Linearity | 2000 IU/L |
Automation | Enabled |
Pack Sizes | 1×20 ml + 1×5 ml 1×40 ml + 1×10 ml |
Kit components | Reagent bottles Instructions for Use. |
Storage Temperature | 2-8 0C |
Shelf life | 18 months |
Kit dimensions | 1×20 ml + 1×5 ml – 85X105X45mm 1×40 ml + 1×10 ml – 85X105X45mm |
Approx. weight of kit | 1×20 ml + 1×5 ml – 50 gms 1×40 ml + 1×10 ml – 92 gm |
Production Capacity | 600 mn ml per annum |