CALCIUM CPC

Intended Use:

Test for estimation of Calcium in serum / plasma using O-CPCmethod.

Calcium CPC Test Procedure Video

Calcium CPC Programming Video

Calcium CPC Test Possible Errors

Features:

  • O-Cresolphthlein Complexone Specific To Calcium
  • Fe And Mg Intereference Eliminated
  • Liquid Stable Reagents With Long Shelf Life
  • Fast Result
  • High Linearity

Specification

Methodology o-CPC
Sample Type Human Serum or Plasma
Reaction Type End Point
Number Of Reagents Double
Primary Wavelength 575 nm
Secondary Wavelength NA
Flow Cell Temperature 37 0C
Blank Settings Reagent
Blank Absorbance Limit < 0.200
Reagent Volume Required Per Test 1000 µl
Sample Volume Required Per Test 10 µL
Incubation Time 5 Minute
Incubation Temperature Room Temperature
Standard Available in the Kit Yes
Standard Concentration 10 gm/dL
Factor NA
Linearity 15 gm/dL
Final Colour Stability 1 hour
Automation Enabled
Pack Sizes 2×25 ML
2×100 ML
Kit components Reagent bottles – 1 bottle each for Reagent and Diluent.
Standard vial.
Instructions for Use
Storage Temperature 2-8 0C
Shelf life 24 months
Kit dimensions 2X25 ml – 85x105x45 mm
2X100 ml – 105x55x145 mm
Approx. weight of kit 2X25 ml – 92 gms
2X100 ml – 185 gms
Production Capacity 600 mn ml per annum