CHLORIDE
Intended Use:
Test for estimation of Chloride in serum / plasma using Mercuric Thiocyanate method.
Chloride Test Procedure
Chloride Programming Video
Possible Errors in Chloride Test and Solutions
Features:
- Mercuric Thiocyanate Specific to Chloride
- Liquid Stable Reagent with Long Shelf Life
- Fast Result
- High Linearity
- Suitable for Fully Automated Chemistry Analyzer
Specification
Methodology | Mercuric Thiocynate |
Reaction Type | End Point |
Reaction Direction | Increasing |
Sample Type | Human Serum or Plasma |
Number Of Reagents | Single |
R1:R2 Mixing Ratio | NA |
Primary Wavelength | 510 nm |
Secondary Wavelength | NA |
Flow Cell Temperature | 37 0C |
Blank Settings | Reagent |
Blank Absorbance Limit | < 0.200 |
Reagent Volume Required Per Test | 1000 µl |
Sample Volume Required Per Test | 10 µL |
Incubation Time | 1 Minutes |
Incubation Temperature | 37 0C |
Standard Available in the Kit | YES |
Standard Concentration | 100 mMol/L |
Factor | NA |
Linearity | 130 mMol/L |
Automation | Enabled |
Pack Sizes | 1×25 ml 2×50 ml |
Kit components | Reagent bottle. Standard vial. Instructions for Use |
Storage Temperature | 2-8 0C |
Shelf life | 24 months |
Kit dimensions | 1×25 ml – 85x105x45 mm 2×50 ml – 105x135x45 mm |
Approx. weight of kit | 1×25 ml – 50 gms 2×50 ml – 185 gms |
Production Capacity | 600 mn ml per annum |