LDL CHOLESTEROL-DIRECT
Intended Use:
Test for estimation of LDL- Cholesterol in serum / plasma using CHOD-POD method.

LDL Cholesterol Direct Test Procedure
LDL Cholesterol Direct Programming Video
Possible Errors in LDL Cholesterol Direct Test and Solutions
Features:
- LDL Cholesterol Specific Detergents Employed
- Liquid Stable Reagents With Long Shelf Life
- Suitable For Fully Automated Chemistry Analyzer
- Fast Result
- High Linearity
Specification
| Methodology | Special Polymer Detergent (CHOD-POD) |
| Reaction Type | End Point |
| Sample Type | Human Serum or Plasma |
| Number Of Reagents | Two |
| R1:R2 Mixing Ratio | 3:1 |
| Primary Wavelength | 546 nm |
| Secondary Wavelength | 630 nm |
| Flow Cell Temperature | 37 0C |
| Blank Settings | Distilled Water |
| Blank Absorbance Limit | <0.300 |
| Reagent Volume Required Per Test | 800 µl (600 µl R1 + 200 µl R2) |
| Sample Volume Required Per Test | 5 µL |
| Incubation Time | 10 Minutes |
| Incubation Temperature | 37 0C |
| Standard Available in the Kit | Yes |
| Standard Concentration | On calibrator vial |
| Factor | NA |
| Linearity | 400 mg/dL |
| Automation | Enabled |
| Pack Sizes | 1×15 ml + 1×5 ml 2×30 ml + 2×10 ml |
| Kit components | Reagent bottles. Calibrator vial. Instructions for Use. |
| Storage Temperature | 2-8 0C |
| Shelf life | 18 months |
| Kit dimensions | 1×15 ml + 1×5 ml – 85X105X45 2×30 ml + 2×10 ml – 85X105X45 |
| Approx. weight of kit | 1×15 ml + 1×5 ml – 50 gms 2×30 ml + 2×10 ml -92 gms |
| Production Capacity | 600 mn ml per annum |