CK-MB

Intended Use:

Test for estimation of CK-MB activity in serum / plasma using UV-kinetic method.

CK-MB Test Procedure

CK_MB Programming Video

Features:

  • Hexokinase / G6PDH – NAC Activated. Immuno Inhibition For M-Subunit
  • Liquid Stable Reagents with Long Shelf Life
  • Fast Result
  • High Linearity
  • LCF (Lipid Clearing Factor) Incorporated for Lipaemic Sample
  • Suitable for Fully Automated Chemistry Analyzer

Specification

Methodology UV Kinetic
Reaction Type Kinetic
Reaction Direction Increasing
Sample Type Human Serum or Plasma
Number Of Reagents Two
R1:R2 Mixing Ratio 4:1
Primary Wavelength 340 nm
Secondary Wavelength NA
Flow Cell Temperature 37 0C
Blank Settings Distilled Water
Blank Absorbance Limit  <0.700
Reagent Volume Required Per Test 1000 µl
Sample Volume Required Per Test 50 µL
Delay Time 300 seconds
Read Time 120 seconds
Incubation Temperature 37 0C
Standard Available in the Kit NO
Standard Concentration NA
Factor 6752
Linearity 2000 IU/L
Automation Enabled
Pack Sizes 2×8 ml + 2×2 ml
2×40 ml + 2×10 ml
Kit components Reagent bottles.
Instructions for Use
Storage Temperature 2-8 0C
Shelf life 18 months
Kit dimensions 2×8 ml + 2×2 ml – 85X105X45 mm
2×40 ml + 2×10 ml – 85X105X45 mm
Approx. weight of kit 2×8 ml + 2×2 ml – 92 gms
2×40 ml + 2×10 ml – 185 gms
Production Capacity 600 mn ml per annum